Handling methods for common problems of Pentax total station

handling methods for common problems of Pentax total station:

finishing: liyunli

(I) electronic bubble; Index difference and 2C correction

1. electronic bubble correction: place the instrument on a tripod or correction table, start the machine and level one direction through electronic bubble

press the laser key ¤ +on/off to start the machine, and "electronic bubble tube adjustment" appears. Press F3 to set zero when the forward and reverse mirrors aim at the same target (approximately 90 °); PRESS ENT after adjustment

remarks: if the deviation of the electronic bubble is very large, the software correction is unsuccessful (repeatedly prompt for alignment and set to zero.) It is a compensator problem, which needs to be sent to the local maintenance department for treatment

2. index difference correction: place the instrument on a tripod or calibration table and level it

f1+f4+on/off: start the machine, select "2-axis" (no need to correct 3-axis), the front and back mirrors aim at the same target, press F3 to set zero, and press ent

after adjustment. Remarks: if the index difference cannot be corrected, the reason is: ① the perpendicularity of the horizontal axis (support difference) exceeds the limit; ② Horizontal axis levelness (2C) exceeds the limit; If it is confirmed that there are no problems with these two indicators, the most important part of f1+f2 experimental machine can be developed as spindle +f4+on/off (select RL) → ent. ③ If it still fails, it is necessary to correct the index difference from no compensation: f1+f4+on/off: ① 2-axis ② 1-axis ③ no, first select "③ no" for correction, then select "② 1-axis" for correction, and finally make correction for 2-axis. If the above methods are not available, it is the problem of the compensator. Please send it to the local maintenance department for treatment

3.2c correction: like other domestic instruments and theodolites, adjust the left and right correction screws of the crosshair of the eyepiece

(II) software problem

1. high input instrument and high prism crash

2. horizontal distance cannot be calculated

3. wrong setting out azimuth

the above software problems can generally be solved by initializing the coordinate data of the instrument. If the initialization cannot be solved, the program needs to be reinstalled. Please send it to the local branch for program upgrade

method: ① when f2+f5+on/off is turned on, the "coordinate data initialization" appears. The practice of Jinan experimental machine we have seen in the world should be called a material experimental machine. PRESS ENT key, and the instrument will automatically return to the electronic bubble interface; ② Press ESC → F5 → F1 → F1 → select the fifth item "clear all" → ent (R300 instrument does not need to do this)

remarks: after the program is upgraded, the instrument must be calibrated in the following steps in sequence:

① factory load: f2+f4+on/off power on, the instrument version number "version number v***" appears. Press F11 to operate the PC control system operating on windows → F5 → ent, and four options appear, Select the default option and press ent. The instrument automatically completes and returns to the electronic bubble interface

② the instrument language is changed to Chinese: ENT → F5 → F1 → F5 → F5 → F4 → ent → F5 → on/off

③ coordinate data initialization: f2+f5+on/off

the following program versions do not need to be upgraded: R300 instrument v34b

r300x instrument v342a

r300x instrument v344. Or v344a

(III) common errors:

① the "angle measuring part error operation cylinder is prone to error 70" motherboard problem, which needs to be sent to the maintenance department for treatment

② the "angle measuring part error 717375" perpendicularity plate or reading head has dust, which needs to be cleaned

③ there is dust on the "angle measuring part error 727476" horizontal dial or reading head, which needs to be cleaned

④ the problem of "ranging Department error 50515253" needs to be sent to Shanghai maintenance department for treatment.

⑤ the problem of "inappropriate state, too strong sunlight" needs to be sent to Shanghai maintenance department for treatment.

⑥ the "target is too close" the measurement target is too close (less than 1 meter), or the optical axis is seriously deviated from

⑦ the "inappropriate state" "target is wrong" the measurement target and measurement mode are wrong, Please select the correct measurement mode

⑧ for "angle measurement data error", please consult the maintenance department

⑨ for "reading error", please consult the maintenance department

⑩ for "P data of EDM", please consult the maintenance department